Line 108: | Line 108: | ||
* Multiple simultaneous users are attempting to read/write to the database at the same time | * Multiple simultaneous users are attempting to read/write to the database at the same time | ||
* If you receive the error " | * If you receive the error " | ||
+ | </ | ||
+ | | 5 | All | Application fails to load computed data: Under the following scenario, the application request and state handling has a known bug that results in computed data not loading within the application until a application close and re-open is performed:< | ||
+ | * A patient/ | ||
+ | * Data begins computing (e.g.: a dose, beam, or device) | ||
+ | * While the data is computing the user navigates to another user interface within the application | ||
+ | * If the data finishes computing for the first user interface while the user is navigated to a new user interface and then the user navigates back to the first user interface, the application may not know the data is computed to display or use the results. | ||
</ | </ | ||
==== DICOM Import ==== | ==== DICOM Import ==== |