 The main high level requirements for using the decimal Launcher and decimal eRT application are: The main high level requirements for using the decimal Launcher and decimal eRT application are:
-  - .decimal Direct (using Auth0) account credentials+  - .decimal Direct account credentials
     - This account needs association with a Site ID (.decimal Customer Account number) that has been licensed for use of the decimal eRT application     - This account needs association with a Site ID (.decimal Customer Account number) that has been licensed for use of the decimal eRT application
     - This account needs specific .decimal Permission for accessing the eRT modules     - This account needs specific .decimal Permission for accessing the eRT modules
Line 46: Line 46:
 The [[electronrt:userguide:tutorials:all_tutorials#organization_configuration| Organization Configuration]] user guide section describes each of the blocks and configuration interfaces. The [[electronrt:userguide:tutorials:all_tutorials#organization_configuration| Organization Configuration]] user guide section describes each of the blocks and configuration interfaces.
  
 +=== DICOM Receiver ===
 +
 +The decimal ElectronRT application supports importing DICOM patients via a DICOM listener/receiver. Refer to the [[electronrt:userguide:dicom_receiver|ElectronRT DICOM Receiver setup guide]] for instructions for enabling and configuring this method of DICOM patient import into the eRT application.
 ==== Keyboard and Mouse Controls ==== ==== Keyboard and Mouse Controls ====
  
Line 51: Line 54:
  
  
-===== Feature/Task Descriptions =====+===== Application Usage =====
  
-The following task descriptions are thorough guides providing complete information about each task within the decimal eRT application.+The following sections are thorough guides providing complete information about each task within the decimal eRT application.
  
   * [[electronrt:userguide:tutorials:all_tutorials#launching_electronrt|Launching decimal eRT]]   * [[electronrt:userguide:tutorials:all_tutorials#launching_electronrt|Launching decimal eRT]]
Line 59: Line 62:
   * [[electronrt:userguide:tutorials:all_tutorials#patient_courses|Patient Courses]]    * [[electronrt:userguide:tutorials:all_tutorials#patient_courses|Patient Courses]] 
   * [[electronrt:userguide:tutorials:all_tutorials#electron_treatment_plans|Electron Plans]]   * [[electronrt:userguide:tutorials:all_tutorials#electron_treatment_plans|Electron Plans]]
 +    * [[electronrt:userguide:tutorials:all_tutorials#prescription|Prescription]]
     * [[electronrt:userguide:tutorials:all_tutorials#patient_geometry|Patient Geometry]]     * [[electronrt:userguide:tutorials:all_tutorials#patient_geometry|Patient Geometry]]
       * [[electronrt:userguide:tutorials:all_tutorials#points|Points]]       * [[electronrt:userguide:tutorials:all_tutorials#points|Points]]
Line 66: Line 70:
     * [[electronrt:userguide:tutorials:all_tutorials#electron_beams|Electron Beams]]     * [[electronrt:userguide:tutorials:all_tutorials#electron_beams|Electron Beams]]
       * [[electronrt:userguide:tutorials:all_tutorials#electron_block_creation|Electron Block Creation]]       * [[electronrt:userguide:tutorials:all_tutorials#electron_block_creation|Electron Block Creation]]
 +      * [[electronrt:userguide:tutorials:all_tutorials#skin_collimator_creation|Electron Skin Collimator Creation]]
       * [[electronrt:userguide:tutorials:all_tutorials#electron_bolus_creation|Electron Bolus Creation]]       * [[electronrt:userguide:tutorials:all_tutorials#electron_bolus_creation|Electron Bolus Creation]]
       * [[electronrt:userguide:tutorials:all_tutorials#intensity_modulator_device|Electron Intensity Modulator]]       * [[electronrt:userguide:tutorials:all_tutorials#intensity_modulator_device|Electron Intensity Modulator]]
 +      * [[electronrt:userguide:tutorials:all_tutorials#electron_beam_normalization|Electron Beam Normalization]]
     * [[electronrt:userguide:tutorials:all_tutorials#plan_approval|Plan Approval]]     * [[electronrt:userguide:tutorials:all_tutorials#plan_approval|Plan Approval]]
     * [[electronrt:userguide:tutorials:all_tutorials#plan_reports|Plan Reports]]     * [[electronrt:userguide:tutorials:all_tutorials#plan_reports|Plan Reports]]
Line 76: Line 82:
   * [[electronrt:userguide:tutorials:all_tutorials#decimal_ert_settings|App Settings]]   * [[electronrt:userguide:tutorials:all_tutorials#decimal_ert_settings|App Settings]]
  
-===== System Usage =====  
  
-==== Improper System Usage ====+ 
 +===== Improper System Usage =====
  
 When using decimal eRT, as with any complex program, there is the potential for misuse. The decimal eRT app is a radiotherapy treatment planning tool that is intended to be used by experienced and knowledgeable professionals working in the field of radiation therapy. When using decimal eRT, as with any complex program, there is the potential for misuse. The decimal eRT app is a radiotherapy treatment planning tool that is intended to be used by experienced and knowledgeable professionals working in the field of radiation therapy.
Line 86: Line 92:
 Below are listed the known application limitations, defects, or inconsistencies. Below are listed the known application limitations, defects, or inconsistencies.
  
-=== General ===+==== General ====
  
-  decimal eRT does not show the MU value for beams and treatment plans. +^ ID  ^ Affected Version(s)  ^  Description   
-  Simultaneous record access:  +|  1  | 0.0.1\\ 0.0.2\\ 0.0.3\\ 0.0.4| decimal eRT does not show the MU value for beams and treatment plans. | 
-    * If multiple instances of the application have the same treatment plan record opened, changes will be prevented from being pushed to the patient database if the treatment plan is changed first by another user. This will cause the other simultaneous instances to become outdated. Changes to the local application on outdated instances will be blocked from pushing plan changes to the patient database, but the changes may still appear to have happened in the local application. Users will be warned of their changes being blocked by the error: ''"Plan revision is outdated. Please close and reopen the treatment plan to continue using this plan."'' + 2  | All                          | Simultaneous record access:<WRAP> 
 +  * If multiple instances of the application have the same treatment plan record opened, changes will be prevented from being pushed to the patient database if the treatment plan is changed first by another user. This will cause the other simultaneous instances to become outdated. Changes to the local application on outdated instances will be blocked from pushing plan changes to the patient database, but the changes may still appear to have happened in the local application. Users will be warned of their changes being blocked by the error: ''Plan revision is outdated. Please close and reopen the treatment plan to continue using this plan."'' 
 +</WRAP>
 +|  3  | All                          | Random Application Crashes:<WRAP> 
 +  * If the [[electronrt:instructions_for_use:instructions_for_use#network_local_data_cache|local or network disk cache]] becomes corrupted the application may crash at seemingly random places while calculations are computing. Clearing the [[electronrt:userguide:tutorials:all_tutorials#local_disk_cache|local disk cache]] within the application settings may help resolve unreproducible application crashes during calculations running. 
 +</WRAP>
 +|  4  | All                          | Network patient database:<WRAP> 
 +  * If the application is set to use a [[electronrt:instructions_for_use:instructions_for_use#patient_database_file_storage|network patient directory]] shared with multiple users there may be a risk for 'database is locked' errors. This is due to the application using SQLite as it's database engine, which does not robustly support concurrent database users and reading/writing. While normal usage between one or two users may not pose a problem with this, having a significant number of users simultaneously accessing the network database or the following scenarios can cause the database to lock and throw this error: 
 +    * The application crashes or a network connection is interrupted during a SQL commit that allows the 'unlock' step to be missed OR 
 +    * The file is locked due to other network usage (e.g.: the file is in use by an external service such as an Antivirus scan, backup, etc) OR 
 +    * Multiple simultaneous users are attempting to read/write to the database at the same time 
 +  * If you receive the error "database is locked" you'll have to wait and try again once the database is unlocked 
 +</WRAP>
 +|  5  | All                          | Application fails to load computed data: Under the following scenario, the application request and state handling has a known bug that results in computed data not loading within the application until a application close and re-open is performed:<WRAP> 
 +  * A patient/plan is opened and a user interface or task is opened 
 +  * Data begins computing (e.g.: a dose, beam, or device) 
 +  * While the data is computing the user navigates to another user interface within the application 
 +  * If the data finishes computing for the first user interface while the user is navigated to a new user interface and then the user navigates back to the first user interface, the application may not know the data is computed to display or use the results. 
 +</WRAP>
 +==== DICOM Import ====
  
-=== Organization Configuration ===+^ ID  ^ Affected Version(s)  ^  Description   ^ 
 +| 1 | All                  | Closing the application immediately after importing a DICOM patient and before the job 'copy_imported_dicom_files_to_course' completes will cause an incomplete CT Image Set when exporting DICOM later in the application. | 
 +| 2 | 0.9.0                | Only plan files with a valid bolus in at least one beam will import properly. |
  
-  - Deleting a CT Curve or Machine can sometimes cause a UI crash. No data is lost and the deletion is still performed. +==== DICOM Export ====
-=== DICOM Import ===+
  
-  - Closing the application immediately after importing a DICOM patient and before the job 'copy_imported_dicom_files_to_course' completes will cause an incomplete CT Image Set when exporting DICOM later in the applcation.+^ ID ^ Affected Version(s)  ^  Description   
 +| 1  | 0.0.1\\ 0.0.2\\ 0.0.3| Exporting local DICOM files and providing no export directory will cause the DICOM files to be exported to the local application installation folder. This will cause the decimal Launcher application to invalidate the local eRT install and require a re-download of the application
 +==== Beams ====
  
-=== Beams === +^ ID ^ Affected Version(s)  ^  Description   ^ 
-  Adding more than one beam with an optimized thickness bolus can sometimes result in the DVH inspection performance being slow. +| 1  | All                  | Adding more than one beam with an optimized thickness bolus can sometimes result in the DVH inspection performance being slow. | 
-  Optimized Bolus outside shape is generated to the target size and not the block size. So if manual edits are performed on the block that expand the block shape, the bolus extents may be undersized. This can be remedied by increasing the bolus block outer border. +| 2  | All                  | Optimized Bolus outside shape is generated to the target size and not the block size. So if manual edits are performed on the block that expand the block shape, the bolus extents may be undersized. This can be remedied by increasing the bolus block outer border. 
-  - Selecting a rind structure as the beam target causes the 'compute_orthogonal_beam_geometry' calculation to take an incredibly long time to determine the best beam approach. This prevents the user from being able to make and configure beam, even manually, while waiting for this calculation to finish.+| 3  | 0.0.3+               | Skin collimator devices may cause interference with fitting inside of the electron bolus when fitting on the patient surface. This is due to sharp corners that result from a rough patient surface contour.|
  
 +==== Plans ====
 +^ ID ^ Affected Version(s)  ^  Description   ^
 +| 1  | All                  | (Approved Plans) Users should understand that the eRT application stores only user input information and not computed data. However, since machine models and dose calculation algorithms can change over time, certain calculated data and equipment model data that is considered critical to the patient's treatment record is stored with each plan upon approval. This stored data includes: treatment device shapes (e.g. bolus, collimators, intensity modulators), 3D dose fields per beam, and beam model data. However, it should be noted that data derived from these stored results may be recomputed and is therefore subject to change if Approved Plans are loaded in future eRT versions. One such example would be DVH curves. While the dose is stored and will never change for an Approved Plan, the sampling scheme used in DVH calculations can change over time, therefore DVH curves may change even for Approved Plans if such changes are made in the eRT application. Patient-to-applicator collisions, structure volumes, and QA Dose Calculations are additional examples of information that could change in future application versions as this data remains computed, not stored, even for Approved Plans. |
 +| 2  | 0.0.1 - 0.0.4               | (Unapproved Plans) Unapproved treatment plans created versions earlier than eRT version 0.0.4 will not have the plan prescription filled in (denoted by the error: ''"Prescription statement not found in plan"'') resulting in dose not being displayed within the treatment plan displays. Users can resolve this by re-adding the prescription that previously existed in the corresponding Course. |
 +| 3  | 0.0.1 - 0.7.0               | (Approved Plans) Approved treatment plans created versions earlier than eRT version 0.8.0 that contain a skin collimator will no longer be able to load the approved skin collimator when opening the plan. Users will be required to clone the approved plan and re-approve the plan to allow the skin collimator devices to be exported or ordered.|
  
  