electronrt:userguide:userguide
Differences
This shows you the differences between two versions of the page.
Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
electronrt:userguide:userguide [2022/06/02 20:17] – [DICOM Import] kerhart | electronrt:userguide:userguide [2023/03/09 20:59] (current) – [General] dpatenaude | ||
---|---|---|---|
Line 34: | Line 34: | ||
The main high level requirements for using the decimal Launcher and decimal eRT application are: | The main high level requirements for using the decimal Launcher and decimal eRT application are: | ||
- | - .decimal Direct | + | - .decimal Direct account credentials |
- This account needs association with a Site ID (.decimal Customer Account number) that has been licensed for use of the decimal eRT application | - This account needs association with a Site ID (.decimal Customer Account number) that has been licensed for use of the decimal eRT application | ||
- This account needs specific .decimal Permission for accessing the eRT modules | - This account needs specific .decimal Permission for accessing the eRT modules | ||
Line 54: | Line 54: | ||
- | ===== Common Planning Tasks ===== | + | ===== Application Usage ===== |
- | The following | + | The following |
* [[electronrt: | * [[electronrt: | ||
Line 82: | Line 82: | ||
* [[electronrt: | * [[electronrt: | ||
- | ===== System Usage ===== | ||
- | ==== Improper System Usage ==== | + | |
+ | ===== Improper System Usage ===== | ||
When using decimal eRT, as with any complex program, there is the potential for misuse. The decimal eRT app is a radiotherapy treatment planning tool that is intended to be used by experienced and knowledgeable professionals working in the field of radiation therapy. | When using decimal eRT, as with any complex program, there is the potential for misuse. The decimal eRT app is a radiotherapy treatment planning tool that is intended to be used by experienced and knowledgeable professionals working in the field of radiation therapy. | ||
Line 101: | Line 101: | ||
| 3 | All | Random Application Crashes:< | | 3 | All | Random Application Crashes:< | ||
* If the [[electronrt: | * If the [[electronrt: | ||
+ | </ | ||
+ | | 4 | All | Network patient database:< | ||
+ | * If the application is set to use a [[electronrt: | ||
+ | * The application crashes or a network connection is interrupted during a SQL commit that allows the ' | ||
+ | * The file is locked due to other network usage (e.g.: the file is in use by an external service such as an Antivirus scan, backup, etc) OR | ||
+ | * Multiple simultaneous users are attempting to read/write to the database at the same time | ||
+ | * If you receive the error " | ||
+ | </ | ||
+ | | 5 | All | Application fails to load computed data: Under the following scenario, the application request and state handling has a known bug that results in computed data not loading within the application until a application close and re-open is performed:< | ||
+ | * A patient/ | ||
+ | * Data begins computing (e.g.: a dose, beam, or device) | ||
+ | * While the data is computing the user navigates to another user interface within the application | ||
+ | * If the data finishes computing for the first user interface while the user is navigated to a new user interface and then the user navigates back to the first user interface, the application may not know the data is computed to display or use the results. | ||
</ | </ | ||
==== DICOM Import ==== | ==== DICOM Import ==== | ||
Line 106: | Line 119: | ||
^ ID ^ Affected Version(s) | ^ ID ^ Affected Version(s) | ||
| 1 | All | Closing the application immediately after importing a DICOM patient and before the job ' | | 1 | All | Closing the application immediately after importing a DICOM patient and before the job ' | ||
- | | 2 | All | + | | 2 | 0.9.0 |
==== DICOM Export ==== | ==== DICOM Export ==== |
electronrt/userguide/userguide.1654201042.txt.gz · Last modified: 2022/06/02 20:17 by kerhart