-===== Walk-through =====+====== Electron Block Walk-through Tutorial ======
  
  This walk-through is intended to be an easy to follow step-by-step guide for completing a "standard clinical electron" set-up. It covers all steps involved in designing and ordering an electron aperture from .decimal. The intent is that after completing this walk-through you will have gained the knowledge and understanding necessary to complete a clinical electron set up utilizing the decimal3D App (it should be noted that the values provided herein are not intended to represent clinical guidelines and should in no way be considered a recommendation for values that are appropriate for clinical use).  This walk-through is intended to be an easy to follow step-by-step guide for completing a "standard clinical electron" set-up. It covers all steps involved in designing and ordering an electron aperture from .decimal. The intent is that after completing this walk-through you will have gained the knowledge and understanding necessary to complete a clinical electron set up utilizing the decimal3D App (it should be noted that the values provided herein are not intended to represent clinical guidelines and should in no way be considered a recommendation for values that are appropriate for clinical use).
Line 40: Line 40:
   - Now you will set the beam entry point by tapping on the patient model at the desired location. Tap again to change the location if needed, then press //Next// to accept this position {{ :decimal3d:instructions_for_use:img_0119.png?400 |}}   - Now you will set the beam entry point by tapping on the patient model at the desired location. Tap again to change the location if needed, then press //Next// to accept this position {{ :decimal3d:instructions_for_use:img_0119.png?400 |}}
   - You are now presented with all the remaining parameters needed to fully define the desired treatment beam.   - You are now presented with all the remaining parameters needed to fully define the desired treatment beam.
-    - These include: SSD, Applicator Size, Gantry Angle, Couch Angle, and Collimator Angle+    - These include: SSD, Applicator Size, Gantry Angle, Couch Angle, and Collimator Angle (note: as of v1.7 SSD now includes a second input that allows SSD setting to 0.1cm resolution)
     - Adjust the parameters as needed to meet the clinical needs of the patient at hand. {{ :decimal3d:instructions_for_use:beamedit0.png?400 |}}     - Adjust the parameters as needed to meet the clinical needs of the patient at hand. {{ :decimal3d:instructions_for_use:beamedit0.png?400 |}}
       - You may also wish to rotate the model to ensure an appropriate treatment angle has been selected (and to check for possible collisions of the applicator with the patient). You can also switch to a Beam's Eye View (BEV) where you can view and measure the block shape (note the BEV shape is projected to the isocenter plane). {{ :decimal3d:instructions_for_use:beamedit1.png?400 |}}       - You may also wish to rotate the model to ensure an appropriate treatment angle has been selected (and to check for possible collisions of the applicator with the patient). You can also switch to a Beam's Eye View (BEV) where you can view and measure the block shape (note the BEV shape is projected to the isocenter plane). {{ :decimal3d:instructions_for_use:beamedit1.png?400 |}}
Line 48: Line 48:
     - Select a shipping priority, review the Shipping and Billing Address information, and press //Order// to securely transmit your order information to .decimal for the cutout to be manufactured. {{ :decimal3d:instructions_for_use:order.png?400 |}}     - Select a shipping priority, review the Shipping and Billing Address information, and press //Order// to securely transmit your order information to .decimal for the cutout to be manufactured. {{ :decimal3d:instructions_for_use:order.png?400 |}}
     - Wait for 30-60 seconds to receive a confirmation of successful order.     - Wait for 30-60 seconds to receive a confirmation of successful order.
-  - Once your order has been placed, the DICOM Export option will become available. Simply click the "Send" icon in the top right corner (next to the Order button) and your treatment device information will be sent to your preconfigured DICOM receiver via a DICOM RT Plan file.+  - Once your order has been placed, the DICOM Export option will become available. Simply click the "Send" icon in the top right corner (see red circle below) and your treatment device information will be sent to your preconfigured DICOM receiver via a DICOM RT Plan file. {{ :decimal3d:instructions_for_use:dicom_send_icon.png?200 |}}
   - If you encounter any difficulties or problems contact our customer service team at 1.800.255.1613   - If you encounter any difficulties or problems contact our customer service team at 1.800.255.1613
  