Line 17: Line 17:
 It is the responsibility that the user performs end-to-end testing prior to the clinical implementation of the decimal eRT app. The user should follow accepted industry guideline (such as AAPM TG244) for the end-to-end testing. This testing should be performed by qualified personnel. It is the responsibility that the user performs end-to-end testing prior to the clinical implementation of the decimal eRT app. The user should follow accepted industry guideline (such as AAPM TG244) for the end-to-end testing. This testing should be performed by qualified personnel.
  
-It is the responsibility of the facility to ensure that all users of the decimal eRT treatment planning system have had training on the product and possess the appropriate clinical education, experience, and (where applicable) licensure to develop clinical treatment plans. This includes, but is not limited to, the application training provided by .decimal staff. +It is the responsibility of the facility to ensure that all users of the decimal eRT treatment planning system have had training on the product and possess the appropriate clinical education, experience, and (where applicable) licensure to develop clinical treatment plans. This includes, but is not limited to, the application training provided by .decimal staff
 + 
 +Before commissioning it is required that an initial Acceptance Testing process is completed. This testing will ensure that the decimal eRT application functions properly within the installed clinical environment of the end users. This testing will utilize manufacturer supplied inputs, including beam models, as the goal is only to ensure proper system operation, not assess the accuracy or quality of user commissioning activities. As such, the Acceptance Testing, is a prerequisite to, not a replacement for commissioning and end-to-end testing. Details regarding Acceptance Testing can be found [[electronrt:instructions_for_use:acceptance_testing|here]].
  
 It is recommended that users follow acceptable global standards during the commissioning of the decimal eRT product. During the clinical set up, the following should be tested to ensure clinical safety prior to treatment: It is recommended that users follow acceptable global standards during the commissioning of the decimal eRT product. During the clinical set up, the following should be tested to ensure clinical safety prior to treatment:
Line 135: Line 137:
 === Internet Data Transfer === === Internet Data Transfer ===
  
-Since electron therapy generally requires frequent use of patient-specific devices that are often not fabricated in-house, there is a necessity to transmit device manufacturing parameters and information to .decimal servers to all for fabrication of the custom devices. In order to protect patient privacy, .decimal's proprietary order file format contains only the minimal data necessary to manufacture each requested device. .decimal values the privacy of patients and security of our customer's sensitive information and we believe the best safeguard to protect critical data is to ensure it doesn’t leave your facility. As such, no PHI, PII, or any sensitive customer billing/payment information is contained in the order files sent to .decimal.+Since electron therapy generally requires frequent use of patient-specific devices that are often not fabricated in-house, there is a necessity to transmit device manufacturing parameters and information to .decimal servers to all for fabrication of the custom devices. In order to protect patient privacy, .decimal's proprietary order file format contains only the minimal data necessary to manufacture each requested device. .decimal values the privacy of patients and security of our customer's sensitive information and we believe the best safeguard to protect critical data is to ensure it doesn’t leave your facility. As such, no PHI, PII, or any sensitive customer billing/payment information is contained in the order files sent to .decimal. Order files will contain the name, business email address, and business phone number of the user placing the order, as well as the date the order is placed.
  
 <imgcaption internet_data_transfer|Internet Data Transfer>{{ :electronrt:instructions_for_use:decimal_ert_internet_transmission.png?direct&400 |}}</imgcaption> <imgcaption internet_data_transfer|Internet Data Transfer>{{ :electronrt:instructions_for_use:decimal_ert_internet_transmission.png?direct&400 |}}</imgcaption>
Line 157: Line 159:
   * All date and time notifications in decimal eRT should match current Windows OS date and time, including proper use of daylight savings time where appropriate (note: decimal eRT will display in 24 hour format, while Windows may display in am/pm depending on local settings)   * All date and time notifications in decimal eRT should match current Windows OS date and time, including proper use of daylight savings time where appropriate (note: decimal eRT will display in 24 hour format, while Windows may display in am/pm depending on local settings)
  
 +FIXME Add figure with the Patient Coordinate system here
  
 ==== Data Validation and Limits ==== ==== Data Validation and Limits ====
Line 341: Line 344:
 ---- ----
  
-<WRAP centeralign>Copyright © 2020 - 2022 .decimal, LLC. All Rights Reserved. \\ +<WRAP centeralign>Copyright © 2020 - 2025 .decimal, LLC. All Rights Reserved. \\ 
 //decimal eRT® is a trademark of .decimal, LLC.// \\ //decimal eRT® is a trademark of .decimal, LLC.// \\
  
 121 Central Park Place, Sanford, FL 32771 \\ 1-800-255-1613</WRAP> 121 Central Park Place, Sanford, FL 32771 \\ 1-800-255-1613</WRAP>
electronrt/instructions_for_use/instructions_for_use.1764339238.txt.gz · Last modified: by kerhart