====== Electron Bolus Creation ====== | ====== Electron Bolus Creation ====== | ||
- | An electron bolus is an option | + | An electron bolus is an optional |
< | < | ||
- | There are four options for boluses in the decimal eRT app as defined below: | + | |
===== Optimized Thickness Bolus ===== | ===== Optimized Thickness Bolus ===== | ||
- | The first option a user has for creating | + | The first option a user has for creating |
< | < | ||
- | If the need arises | + | For complex and challenging plans, |
- | In the advanced options you can add the following | + | In the advanced options you can view and manage various |
- | The following operators are only available when first creating a new bolus. | + | **Construction Operators (only available when first creating a new bolus):** |
- | * **Automated Marching** | + | * **Automated Marching** |
+ | | ||
* **Geometric Sequence** | * **Geometric Sequence** | ||
+ | * This selection performs the following sequence of operators: Create, Smooth, Truncate. | ||
* **Single shift Sequence** | * **Single shift Sequence** | ||
+ | * This selection performs the following sequence of operators: Create, Smooth, Isodose Shift, Smooth, Truncate. | ||
* **Double Shift Sequence** | * **Double Shift Sequence** | ||
+ | * This selection performs the following sequence of operators: Create, Smooth, Isodose Shift, Smooth, Isodose Shift, Smooth, Truncate. | ||
* **Create** | * **Create** | ||
+ | * This operator constructs the proximal bolus surface by computing the difference between the physical depth to the distal PTV surface and the depth of the Planning Isodose level along each ray line of the bolus construction grid (for ray lines not intersecting the PTV, an extension is performed using the heights of the points that do intersect, which helps ensure the PTV edges are adequately covered). | ||
- | FIXME Kevin may want to add details here for what each operator does in terminology that best jives with end uers. | + | **Modification Operators (only available after a construction operator has been performed):** |
- | + | ||
- | The following operators are only available after one of the above bolus creation operators have been performed: | + | |
* **Smooth** | * **Smooth** | ||
+ | * This operator performs a Gaussian type smoothing on the current bolus proximal surface, which uses a distance weighted average of the neighboring point heights to compute a new height for each bolus surface point. | ||
* **Isodose shift** | * **Isodose shift** | ||
+ | * This operator updates the bolus proximal surface by using the dose results with the current bolus in place to adjust the bolus heights by the difference between the position of the Planning Isodose contour and the position of the distal PTV surface. | ||
* **Truncate** | * **Truncate** | ||
+ | * This operator updates a bolus proximal surface by reducing the height of the bolus in the region outside the Block Outer Border. The height is reduced to be slightly above the highest point in the modulated region. It should be mentioned that truncating a bolus may limit the ability to shift points upward (increase in thickness) when performing manual isodose shift operators after the truncate. | ||
* **Specified Shift** | * **Specified Shift** | ||
+ | * This operator updates a bolus proximal surface by shifting all points (even the region beyond the Block Outer Border) by the specified distance. | ||
* **Intensity Modulation** | * **Intensity Modulation** | ||
+ | * This operator manually adds an [[electronrt: | ||
- | For each operator | + | If an operator is selected from the list the UI will update |
- | Here we have an example of adding the automated marching to a new bolus: | + | Below is an example of adding the automated marching to a new bolus: |
< | < | ||
Line 42: | Line 50: | ||
< | < | ||
- | | + | The advanced optimized thickness bolus design parameters allow the user to fine tune the bolus design for specific operators. These options include: |
+ | * **Block Outer Border:** | ||
+ | * The distance to extend the bolus surface beyond the block/ | ||
+ | * **Target Inner Border:** | ||
+ | * Specifies the distance to offset the target projection inward when defining the region over which initial construction and smoothing operators act. This is used to reduce the effects that the rapid PTV height changes near the target edges can have on the construction and smoothing operators. | ||
+ | * **Depth Beyond Target:** | ||
+ | * The beams-eye-view distance past the selected target (along the CAX direction) at which the patient-side surface will be truncated. | ||
+ | * **Minimum Thickness: | ||
+ | * The minimum allowable thickness of the bolus as measured along a beam aligned rayline. | ||
+ | * **Distal PTV Smoothing: | ||
+ | * The smoothing ball radius used to smooth the target structure. A larger radius will result in more smoothing but will also cause a greater deviation of the target from the original structure. Note that setting this value to 0 mm will result in no smoothing for the target structure. | ||
+ | * **Smoothing Weight:** (Smoothing operator only) | ||
+ | * The exponential smoothing weight coefficient (smaller value increases smoothing) | ||
+ | * **Smoothing Size:** (Smoothing operator only) | ||
+ | * The smoothing distance factor (larger values spread the smoothing range further) | ||
+ | * **Additional Thickness: | ||
+ | * The thickness to add to the bolus proximal surface (use a negative value to reduce height) | ||
+ | |||
+ | Additionally by selecting the "Bolus Design History" | ||
< | < | ||
+ | When designing an optimized thickness bolus, you also have the option to select the patient external structure. By default, this is set to the external structure defined in the course. | ||
+ | |||
+ | < | ||
+ | |||
+ | If you want to make changes to the beam without having the optimized thickness bolus re-calculate, | ||
+ | |||
+ | < | ||
===== Uniform Thickness Bolus ===== | ===== Uniform Thickness Bolus ===== | ||
- | When adding a uniform thickness bolus the user must add the UTB to all the beams in the plan. Once the plan level bolus has been selected | + | Uniform Thickness Bolus devices are designed at the plan level, but they must be added to each intended beam individually to ensure |
< | < | ||
- | ===== Bolus as a Structure ===== | + | ===== Structure |
- | The user also has the option to instead include | + | The user also has the option to use existing plan structure as a bolus. Once this option is selected, |
< | < | ||
Line 60: | Line 93: | ||
< | < | ||
- | Once the structure is selected the bolus will be added to this beam and dose re-calculated | + | Once the structure is selected the bolus will be added to this beam and dose re-calculated. |
**Note:** The bolus is displayed in the BEV for this beam even though the " | **Note:** The bolus is displayed in the BEV for this beam even though the " | ||
Line 66: | Line 99: | ||
< | < | ||
- | If you with to import a bolus that is not currently in your structure list please refer to the [[electronrt: | + | If you wish to import a bolus that is not currently in your structure list please refer to the [[electronrt: |
- | ===== IMET Device ===== | + | |
- | Additionally if any type of bolus is added to a beam the user is able to also include an intensity | + | ===== Intensity Modulator Device ===== |
+ | |||
+ | Additionally if any type of bolus is added to a beam the user is able to also include an Intensity | ||
**NOTE:** In order to add an IMET device the selected beam MUST have a valid bolus. | **NOTE:** In order to add an IMET device the selected beam MUST have a valid bolus. | ||
+ | <WRAP center round important 60%> | ||
+ | Circular applicator shapes (e.g.: Siemens 4x4) are not supported for intensity modulators. | ||
+ | </ | ||
+ | |||
+ | |||
+ | < | ||
< | < | ||
- | And as with a bolus the device can be hidden and un-hidden from the display using the beam controls on the right hand size. | + | As with a bolus, the IM device can be hidden and un-hidden from the display using the beam controls on the right hand size. |
< | < | ||
+ | Users with [[electronrt: | ||
+ | |||
+ | An example file is: | ||
+ | <code csv im_file_example.csv> | ||
+ | 0.3, | ||
+ | -0.3, | ||
+ | </ | ||
+ | |||
+ | < | ||
electronrt/userguide/tutorials/electron_bolus.1600440741.txt.gz · Last modified: (external edit)