electronrt:userguide:tutorials:electron_bolus
Differences
This shows you the differences between two versions of the page.
Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
electronrt:userguide:tutorials:electron_bolus [2020/12/04 17:34] – [Bolus as a Structure] kerhart | electronrt:userguide:tutorials:electron_bolus [2022/05/05 14:39] (current) – [Intensity Modulator Device] dpatenaude | ||
---|---|---|---|
Line 68: | Line 68: | ||
* The thickness to add to the bolus proximal surface (use a negative value to reduce height) | * The thickness to add to the bolus proximal surface (use a negative value to reduce height) | ||
- | Additionally by selecting the "Bolus Design History" | + | Additionally by selecting the "Bolus Design History" |
< | < | ||
+ | When designing an optimized thickness bolus, you also have the option to select the patient external structure. By default, this is set to the external structure defined in the course. | ||
+ | |||
+ | < | ||
+ | |||
+ | If you want to make changes to the beam without having the optimized thickness bolus re-calculate, | ||
+ | |||
+ | < | ||
===== Uniform Thickness Bolus ===== | ===== Uniform Thickness Bolus ===== | ||
Line 93: | Line 100: | ||
If you wish to import a bolus that is not currently in your structure list please refer to the [[electronrt: | If you wish to import a bolus that is not currently in your structure list please refer to the [[electronrt: | ||
+ | |||
===== Intensity Modulator Device ===== | ===== Intensity Modulator Device ===== | ||
- | Additionally if any type of bolus is added to a beam the user is able to also include an intensity | + | Additionally if any type of bolus is added to a beam the user is able to also include an Intensity |
**NOTE:** In order to add an IMET device the selected beam MUST have a valid bolus. | **NOTE:** In order to add an IMET device the selected beam MUST have a valid bolus. | ||
+ | <WRAP center round important 60%> | ||
+ | Circular applicator shapes (e.g.: Siemens 4x4) are not supported for intensity modulators. | ||
+ | </ | ||
+ | |||
+ | |||
+ | < | ||
< | < | ||
- | FIXME get a better IM image once that actually works... | ||
- | And as with a bolus the device can be hidden and un-hidden from the display using the beam controls on the right hand size. | + | As with a bolus, the IM device can be hidden and un-hidden from the display using the beam controls on the right hand size. |
< | < | ||
+ | Users with [[electronrt: | ||
+ | |||
+ | An example file is: | ||
+ | <code csv im_file_example.csv> | ||
+ | 0.3, | ||
+ | -0.3, | ||
+ | </ | ||
+ | |||
+ | < | ||
electronrt/userguide/tutorials/electron_bolus.1607103275.txt.gz · Last modified: (external edit)