electronrt:userguide:tutorials:plan_qa
Differences
This shows you the differences between two versions of the page.
Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
electronrt:userguide:tutorials:plan_qa [2020/08/31 17:53] – [IMET-391] User Guide Plan QA anegron | electronrt:userguide:tutorials:plan_qa [2021/09/09 19:17] (current) – [Dose QA] anegron | ||
---|---|---|---|
Line 2: | Line 2: | ||
The ElectronRT App includes a set of QA Tools that allow the user to measure each treatment beam dose on a water phantom, compute dose on bolus simulation CT images and compare results, and export QA PDF reports. The QA Tools block is only enabled for approved treatment plans. | The ElectronRT App includes a set of QA Tools that allow the user to measure each treatment beam dose on a water phantom, compute dose on bolus simulation CT images and compare results, and export QA PDF reports. The QA Tools block is only enabled for approved treatment plans. | ||
===== Dose QA ===== | ===== Dose QA ===== | ||
- | The Dose QA block computes and displays the dose for each beam in the treatment plan to a water phantom. For each beam, the user is able to define the SSD and the depth of a measurement plane. This measurement plane will be used to take a cross-sectional sample of the water phantom dose. The dimensions of the water phantom, as well as the default measurement plane depth and dose grid spacing, are defined and editable in the site configuration. FIXME <Add Link> After defining the measurement plane depth for at least one beam, the user will be able to click the Export QA Dose Files button to export a QA Dose DICOM file for every measurement plane dose to a local directory. | + | The Dose QA block computes and displays the dose for each beam in the treatment plan to a water phantom. For each beam, the user is able to define the SSD and the depth of a measurement plane. This measurement plane will be used to take a cross-sectional sample of the water phantom dose. If a custom water phantom was added to the [[: |
- | < | + | After defining the measurement plane depth for at least one beam, the user will be able to click the Export QA Dose Files button to export a QA Dose DICOM file for every measurement plane dose to a local directory. The exported DICOM will include: |
+ | - A 1cm thick dose image around the set measurement plane | ||
+ | - [optional] The full 3D water phantom dose - if the user has selected the full 3D dose export option | ||
+ | |||
+ | |||
+ | < | ||
+ | < | ||
===== Bolus QA ===== | ===== Bolus QA ===== | ||
- | The Bolus QA block allows the user to import DICOM CT images and an optional structure set used for bolus simulation | + | The Bolus QA block allows the user to import DICOM CT images and an optional structure set used for bolus QA simulation. The process is very similar to the [[electronrt: |
< | < | ||
- | Dose is computed | + | QA is performed by computing dose on the re-imported CT images |
+ | |||
+ | The comparison view shows the three images simultaneously on the corresponding CT images as well as a graph displaying the DVHs for the original and re-computed dose images. The user can switch between Transverse, Sagittal, and Coronal for the comparison view in the Bolus QA Display Options in the right side controls. | ||
- | The comparison view shows the three images simultaneously on the corresponding CT images as well as a graph displaying the DVHs for the original | + | Users can also set the gamma calculation parameters and required passing rates. This includes: |
+ | * **Passing Rate (%):** The percentage of voxels that must be within tolerance in order for QA to " | ||
+ | * **Spatial Tolerance (cm):** The maximum allowable distance-to-agreement (DTA) value, where DTA is the distance from a point in the original | ||
+ | * **Value Tolerance (%): | ||
+ | * **Max Threshold (%):** The relative dose value below which dose comparison calculations will no longer be performed | ||
< | < | ||
Line 27: | Line 39: | ||
===== Block QA Reports ===== | ===== Block QA Reports ===== | ||
- | The Block QA Reports block allows the user to export PDF printouts of the electron blocks used in the treatment plan. The user can choose to display each block as a physical (true size) or isocenter projection. There is also a printer-friendly option that removes the color from the block projection. | + | The Block QA Reports block allows the user to export PDF printouts of the electron blocks used in the treatment plan. The user can choose to display each block as a physical (true size) or isocenter projection. There is also a printer-friendly option that removes the color from the block projection |
< | < | ||
< | < |
electronrt/userguide/tutorials/plan_qa.1598896431.txt.gz · Last modified: (external edit)